IEC 60601-1 Power Supply 225W
- 4 x 2 x 1 Inches Form factor
- 225 Watts with Forced Air Cooling
- High efficiency up to 94%
- -40 to 70 °C degree operating temperature*
- 12V / 0.5A Fan Output, Thermal Shut-down protection
- 3.37m Hours, Telcordia -SR332-issue 3 MTBF
- Standby Power < 0.5W
Detailed description
Electrical Specifications | |
---|---|
Input Voltage | 85-264 VAC/390 VDC, Universal (Derate from 100% at 100V AC to 95% at 85V AC) |
Input Frequency | 47 - 63 Hz |
Input Current | 115 VAC: 2.2 A max. 230 VAC: 1.1 A max. |
Standby Power | less than 0.5W typical |
Inrush Current | 115 VAC – 25 A, 230 VAC – 45 A, 264 VAC – 75 |
Leakage Current | 300 uA Typical, (N.A. For Class II Option) Touch current <100uA |
Efficiency | 94%(48V), 93%(24V,30V), 92%(12V,15V) |
Hold-up Time | at 225W:10 ms ; 110W: 16 ms |
Power Factor | exceeds 0.95 with Full Load |
Output Power | 225W with 13 CFM, upto 120W Convection |
Line Regulation | +/-0.5% |
Load Regulation | +/-0.5% |
Transient Response | 25% step load change, at 0.1A/uS slew rate, 50% duty cycle, 50Hz=4%, recovery time <5 ms |
recovery time | 55ms typical |
Set Point Tolerance | +/-1 |
Over Current Protection | >110% |
Over Voltage Protection | 110 to 140% |
Short Circuit Protection | Hiccup mode |
Switching Frequency | PFC – 70 to 130 KHz ,PWM – 50-80 KHz |
Operating Temperature | - 40 to + 70°C |
Storage Temperature | - 40 to + 85°C |
Relative Humidity | 5% to 95%, noncondensing |
Altitude | Operating:16,000 ft.; Nonoperating: 40,000 ft. |
MTBF | 3.37m Hours, Telcordia -SR332-issue 3 |
Isolation Voltage | Input to Output – 3000V AC for ITE application Input to GND - 1500 VAC (Not Applicable For Class II Option) |
Cooling | 225W with 13 CFM forced air cooling 6 (refer Mechanical Drawing) up to 120 W with natural convection cooling 6 (refer Derating Curve) |
Model Number | Description | Voltage | Max. Load (Convection) | Max. Load (375 LFM) | Ripple |
---|---|---|---|---|---|
FSP225-1K20M1 | with Screw Terminal | 12 V | 8.33A | 0.0 A | 1% |
FSP225-1K20M1-II | with Screw Terminal | 12 V | 8.33A | 0.0 A | 1% |
FSP225-1K21M1 | with Molex Connector | 15 V | 8.33A | 0.0 A | 1% |
FSP225-1K21M1-II | with Molex Connector | 15 V | 8.33A | 0.0 A | 1% |
FSP225-1K30M1 | with Screw Terminal | 24 V | 6.66A | 0.0 A | 1% |
FSP225-1K30M1-II | with Screw Terminal | 24 V | 6.66A | 0.0 A | 1% |
FSP225-1K31M1 | with Molex Connector | 48 V | 6.66A | 0.0 A | 1% |
FSP225-1K31M1-II | with Molex Connector | 48 V | 6.66A | 0.0 A | 1% |
FSP225-1K40M1 | with Screw Terminal | 30 V | 4.16A | 0.0 A | 1% |
FSP225-1K40M1-II | with Screw Terminal | 30 V | 4.16A | 0.0 A | 1% |
FSP225-1K41M1 | with Molex Connector | 58 V | 4.16A | 0.0 A | 1% |
FSP225-1K80M1 | with Molex Connector | 58 V | 1.56A | 0.0 A | 1% |
FSP225-CK | metal cover kit accessory |
NOTES
- Ripple is peak to peak with 20 MHz bandwidth and 10 µF (Tantalum capacitor) in parallel with a 0.1 µF capacitor at rated line voltage and load ranges.
- Class II means without input Earth pin.
- Combined output power of main output, fan supply shall not exceed the max. Power rating.
- Fan supply output voltage tolerance including set point accuracy, line, and load regulation is +/-10% and Ripple and noise is less than 10%.
- Specifications are for nominal input voltage, 25°C unless otherwise stated.
- 225W with 13CFM forced air cooling and 120W with natural convection cooling at 100 to 264VAC.
- -40 to 0°C startup is guaranteed with spec deviation in output ripple and voltage regulation.
Datasheet
IEC 60601-1 225W