IEC 60601-1 Power Supply 120W

  • 3” x 2” foot print
  • Height 1” above PCB
  • 120 Watts with Forced Air Cooling
  • Approval to EN60601 3 rd Edition, Dual fusing
  • Efficiency upto 93%
  • -40 to 70°C operating temperature, Thermal Shut-down protection
  • Suitable for BF applications
  • Means of Protection : 2xMOPP
  • >3.00m Hours, Telcordia -SR332-issue 3
  • Standby Power < 0.3W
  • Class II option available

Detailed description

Electrical Specifications
Input Voltage85-264 VAC/390 VDC 5 , Universal (see derating under output power)
Input Frequency47 - 63 Hz
Input Current115 VAC – 25 A, 230 VAC – 45 A, 264 VAC – 75 A
Standby Powerless than 0.3W typical
Inrush Current115 VAC – 25 A, 230 VAC – 45 A, 264 VAC – 75 
Leakage Current300 uA Typical, (N.A. For Class II Option) Touch current <100uA
Efficiency93%(48V,58V), 91%(24V,30V), 90%(12V,15V)
Hold-up Time>10 ms typical
Power Factorexceeds 0.95 with Full Load, Active PFC
Output PowerForced cooling : 120W with 300LFM (refer mechanical dimension)
Convection cooling : 100W (for input 100-264 VAC)
(de-rate linearly to 80W @ 85VAC)
Output Voltage Adjustability+/-3%
Line Regulation+/-0.5%
Load Regulation+/-1%
Transient Response25% step load change, at 0.1A/uS slew rate, 50% duty cycle, 50Hz=4% ,recovery time < 5 ms
recovery time55ms typical
Set Point Tolerance+/-1
Over Current Protection>110%
Over Voltage Protection110 to 140%, Latch type (AC recycling required)
Short Circuit ProtectionHiccup mode
Switching Frequency60 KHz
Operating TemperaturePFC – 70 to 130 KHz, PWM – 50-80 KHz
Operating Temperature- 40 to + 70°C
Storage Temperature- 40 to + 85°C
Relative Humidity5% to 95%, noncondensing
AltitudeOperating:16,000 ft.; Nonoperating: 40,000 ft.
MTBF>3.00m Hours, Telcordia -SR332-issue 3
Isolation VoltageInput to Output – 4000 VAC medical applications
Input to GND - 1500 VAC (Not Applicable For Class II Option)
Output to GND- 1500VAC for type BF , 500 VAC for type B (Not Applicable For Class II Option)
CoolingPrimary to Secondary: 2 MOPP, Primary to Earth: 1 MOPP, Secondary to Earth: 1 MOPP

Model NumberDescriptionVoltageMax. Load
(Convection)
Max. Load
(300 LFM)
Ripple
FSP120-1K20M1with Screw Terminal12 V8.33A10.0 A1%
FSP120-1K20M1-IIwith Screw Terminal12 V8.33A10.0 A1%
FSP120-1K21M1with Molex Connector12 V8.33A10.0 A1%
FSP120-1K21M1-IIwith Molex Connector12 V8.33A10.0 A1%
FSP120-1K30M1with Screw Terminal15 V6.66A8.0 A1%
FSP120-1K30M1-IIwith Screw Terminal15 V6.66A8.0 A1%
FSP120-1K31M1with Molex Connector15 V6.66A8.0 A1%
FSP120-1K31M1-IIwith Molex Connector15 V6.66A8.0 A1%
FSP120-1K40M1with Screw Terminal24 V4.16A5.0 A1%
FSP120-1K40M1-IIwith Screw Terminal24 V4.16A5.0 A1%
FSP120-1K41M1with Molex Connector24 V4.16A5.0 A1%
FSP120-1K80M1with Molex Connector24 V4.16A5.0 A1%
FSP120-1K80M1-IIwith Screw Terminal48 V2.08A2.5A1%
FSP120-1K81M1with Screw Terminal48 V2.08A2.5A1%
FSP120-1K81M1-IIwith Molex Connector48 V2.08A2.5A1%
FSP120-1K50M1with Molex Connector48 V2.08A2.5A1%
FSP120-1K50M1-IIwith Screw Terminal30 V3.33A4.0 A1%
FSP120-1K51M1with Screw Terminal30 V3.33A4.0 A1%
FSP120-1K51M1-IIwith Molex Connector30 V3.33A4.0 A1%
FSP120-1K70M1with Molex Connector30 V3.33A4.0 A1%
FSP120-1K70M1-IIwith Screw Terminal58 V1.72A2.07A1%
FSP120-1K71M1with Screw Terminal58 V1.72A2.07A1%
FSP120-1K71M1-IIwith Molex Connector58 V1.72A2.07A1%
FSP120-CKmetal cover kit accessory

NOTES

  1. Ripple is peak to peak with 20 MHz bandwidth and 10 µF (Tantalum capacitor) in parallel with a 0.1 µF capacitor at rated line voltage and load ranges.
  2. Class II means without input Earth connection.
  3. Specifications are for nominal input voltage, 25°C unless otherwise stated.
  4. -40 to 0°C startup is guaranteed with spec deviation in output ripple and voltage regulation.
  5. Functional, not approved.


Datasheet
IEC 60601-1 120W