IEC 60601-1 POWER SUPPLY 75W

    • 3 x 2 x 1 Inches Form factor
    • 75 Watts Convection
    • Approval to EN60601 3 rd Edition, Dual fusing
    • Efficiency upto 93%
    • -40 to 70°C operating temperature, Thermal Shut-down protection
    • Suitable for BF applications
    • Means of Protection : 2xMOPP
    • 2.00m Hours, Telcordia -SR332-issue 3
    • Standby Power < 0.3W
    • Class II option available

    Detailed Description

    Electrical Specifications
    Input Voltage85-264 VAC/390 VDC 5 , Universal (Derate from 75W at 100V AC to 65W at 85V AC)
    Input Frequency47 - 63 Hz
    Input Current115Vac:1.0A(max.), 230Vac:0.5A(max.)
    Standby Powerless than 0.3W typical
    Inrush Current115 VAC – 25 A, 230 VAC – 45 A, 264 VAC – 75 
    Leakage Current300 uA Typical, (N.A. For Class II Option) Touch current <100uA
    Efficiency93%(48V,58V),91%(24V,30V),90%(12V,15V)
    Hold-up Time>16 ms typical
    Power Factorexceeds 0.95 with Full Load, Active PFC
    Output Power75W Convection
    Output Voltage Adjustability+/-3%
    Line Regulation+/-0.5%
    Load Regulation+/-1%
    Transient Response25% step load change, at 0.1A/uS slew rate, 50% duty cycle, 50Hz=4%, recovery time < 5 ms
    recovery time55ms typical
    Set Point Tolerance+/-1
    Over Current Protection>110%
    Over Voltage Protection110 to 140%, Latch type (AC recycling required)
    Short Circuit ProtectionHiccup mode
    Switching Frequency60 KHz typical
    Operating Temperature- 40 to + 70°C
    Storage Temperature- 40 to + 85°C
    Relative Humidity5% to 95%, noncondensing
    AltitudeOperating:16,000 ft.; Nonoperating: 40,000 ft.
    MTBF2.00m Hours, Telcordia -SR332-issue 3
    Isolation VoltageInput to Output – 4000 VAC medical applications.
    Input to GND - 1500 VAC (Not Applicable For Class II Option)
    Output to GND- 1500VAC for type BF , 500 VAC for type B (Not Applicable For Class II Option)
    Protection LevelPrimary to Secondary: 2 MOPP, Primary to Earth: 1 MOPP, Secondary to Earth: 1 MOPP
    Cooling75W with natural convection cooling at 100 to 264VAC.

    Model NumberDescriptionVoltageMax. Load
    (Convection)
    Max. Load
    (375 LFM)
    Ripple
    FSP075-1K20M1with Screw Terminal12 V6.25A0.0 A1%
    FSP075-1K20M1-IIwith Screw Terminal12 V6.25A0.0 A1%
    FSP075-1K21M1with Molex Connector12 V6.25A0.0 A1%
    FSP075-1K21M1-IIwith Molex Connector12 V6.25A0.0 A1%
    FSP075-1K30M1with Screw Terminal15 V5.00A0.0 A1%
    FSP075-1K30M1-IIwith Screw Terminal15 V5.00A0.0 A1%
    FSP075-1K31M1with Molex Connector15 V5.00A0.0 A1%
    FSP075-1K31M1-IIwith Molex Connector15 V3.12A0.0 A1%
    FSP075-1K40M1with Screw Terminal24 V3.12A0.0 A1%
    FSP075-1K40M1-IIwith Screw Terminal24 V3.12A0.0 A1%
    FSP075-1K41M1with Molex Connector24 V3.12A0.0 A1%
    FSP075-1K80M1with Molex Connector24 V1.56A0.0 A1%
    FSP075-1K80M1-IIwith Screw Terminal48 V1.56A0.0 A1%
    FSP075-1K81M1with Screw Terminal48 V1.56A0.0 A1%
    FSP075-1K81M1-IIwith Molex Connector48 V1.56A0.0 A1%
    FSP075-1K50M1with Molex Connector48 V2.50A0.0 A1%
    FSP075-1K50M1-IIwith Screw Terminal30 V2.50A0.0 A1%
    FSP075-1K51M1with Screw Terminal30 V2.50A0.0 A1%
    FSP075-1K51M1-IIwith Molex Connector30 V2.50A0.0 A1%
    FSP075-1K70M1with Molex Connector30 V1.29A0.0 A1%
    FSP075-1K70M1-IIwith Screw Terminal58 V1.29A0.0 A1%
    FSP075-1K71M1with Screw Terminal58 V1.29A0.0 A1%
    FSP075-1K71M1-IIwith Molex Connector58 V1.29A0.0 A1%
    FSP075-CKmetal cover kit accessory

    NOTES

    1. Ripple is peak to peak with 20 MHz bandwidth and 10 µF (Tantalum capacitor) in parallel with a 0.1 µF capacitor at rated line voltage and load ranges.
    2. Combined output power of main output, fan supply shall not exceed max. Power rating.
    3. Fan supply output voltage tolerance including set point accuracy, line and load regulation is +/-10% and Ripple and noise is less than 10%.
    4. Specifications are for nominal input voltage, 25°C unless otherwise stated.
    5. Thermal shutdown feature : The power supply goes in hiccup mode when the temperature of PCB exceeds 110 °C (+/-10 °C).
    6. Output ripple can be more than 10% of the output voltage.


    Datasheet
    IEC 60601-1 75W